ECG-100G Digital Single Channel ECG EKG Machine Electrocardiograph
â– High-resolution thermal printing
array system
â– High resolution digital filter inhibits
baseline drift without affecting ECG
waveform
â– One-touch operation
| Input Circuit | Floating; Protection against Defibrillator effect |
| Lead | Standard 12 Leads, Lead change automatically |
| Patient Current Leakage | <10μA |
| Input Impedance | ≥10ΜΩ |
| Calibrating Voltage | 1mV±3% |
| A/D Conversion | 12 bit |
| Frequency Response | 0.05Hz~150Hz (IEC) |
| Time Constant | >3.2seconds |
|
CMRR
|
>80dB
>100dB (with filter)
|
| EMG Filter | 35Hz (-3dB) / 25Hz (-3dB) |
| Sensitivity | 1/2, 1, 2 (cm/mV), Conversion Deviation≤5% |
| Recording System | Thermal-array
8 dots/mm ( vertical )
16 dots/mm ( horizontal, 25mm/sec )
|
| Paper Speed | 25, 50mm/s±3% |
| Recording Paper | 50mm,20m, high-speed roll |
| Input Circuit Current | ≤50nA |
| Safety Standard | IEC class I, type CF |
| Power Supply |
AC: 220V / 110V, 50/60Hz
DC: 12V rechargeable battery
|
| Skin Voltage Tolerance | ±300mV |
| Noise Level | <15uVp-p |
| Dimensions | 310mm×230mm×70mm |
| Dimension of Packing | 400mm×280mm×160mm |
| Weight | 3.5kg |
Attention please, item could only be shipped out via EMS.
For orders from any other countries where there's no EMS
service or orders shipped by other Express company(like
UPS,DHL, FEDEX,)please email us when you placed your order.
Thanks for your reading and cooperation.
Tell us what you think about this item, share your opinion with other people. Please make sure that your review focus on this item. All the reviews are moderated and will be reviewed within two business days. Inappropriate reviews will not be posted.
Have any question or inquire for this item? Please contact Customer Service. (Our customer representative will get back shortly.)